Exchanging The Fan - Agilent Technologies 1100 Series Reference Manual

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Variable Wavelength Detector
2 Turn off the detector, then on again. Then, restart the Agilent ChemStation. If the
3 After restart, the serial number you have just entered can be seen under the

Exchanging the Fan

When required
Tools required
Parts required
The fan must be installed in the correct orientation to ensure optimum cooling and
operation of the detector.
CAUTION
Preparations for this procedure:
Turn off the lamp.
Switch off the detector, and disconnect the cables.
Remove detector from stack and place it on the
working bench.
Remove the front cover, top cover and top foam
section, see
"Removing the Top Cover and Top
Foams"
on page 103.
106 1100 Series Variable Wavelength Detector Reference Manual
serial number you have just entered is different than the original module serial
number, you will be given the opportunity to edit the configure 1100 access screen
during the restart of the Agilent ChemStation.
Instrument menu of the main user interface screen. The serial number of the detector
can also be seen by typing the following command into the command line:
print sendmodule$ (lvwd, "ser?")
The reply line will give the module serial number.
If the fan is defective or
noisy
Screwdriver POZI 1 PT3
Fan G1314-65004