Performance Specifications - Agilent Technologies 1100 Series Reference Manual

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

8
Specifications

Performance Specifications

Table 62 Performance Specifications Agilent 1100 Series Variable Wavelength Detector
Type
Detection type
Light source
Wavelength range
Short term noise
( ASTM )
Drift
Linearity
Wavelength accuracy
Band width
Specification
Double-beam photometer
Deuterium lamp
190–600 nm
-5
± 0.75 × 10
AU at 254 nm
-4
3 × 10
AU/hr at 254 nm
> 2 AU (5%) upper limit
± 1 nm
6.5 nm typical
Agilent Technologies
Comments
See NOTE on
page
See NOTE on
page 243
See NOTE on
page 243
Self-calibration with deuterium
lines, verification with holmium
oxide filter
241
243.