Exchanging The Leak Sensor - Agilent Technologies 1100 Series Reference Manual

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Variable Wavelength Detector

Exchanging the Leak Sensor

When required If defective
Tools required Screwdriver POZI 1 PT3 Parts
required
1 Turn off the lamp.
2 Switch off the detector, and disconnect the cables and capillaries.
3 Remove the detector from the stack and place it on the working bench.
4 Remove the front cover, top cover and top foam section, see
5 Remove the optical unit, see
116 1100 Series Variable Wavelength Detector Reference Manual
Leak sensor assembly 5061-3356
Cover and Top Foams"
on page 103.
"Removing the Optical Unit"
"Removing the Top
on page 120.