Agilent Technologies 1100 Series Reference Manual page 65

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Figure 10
Holmium Test
When to do the Test
• after recalibration,
• as part of the Operational Qualification/Performance Verification procedure, or
• after flow cell maintenance or repair.
Interpreting the Results
The test is passed successfully when all three wavelengths are within ± 1 nm of the
expected value. This indicates the detector is calibrated correctly.
The test results are currently available on the Agilent ChemStation only.
Holmium Oxide Test Failed
1100 Series Variable Wavelength Detector Reference Manual 63
Troubleshooting and Test Functions
3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents