4 Select Signal 2 - Raw Reference.
5 Observe the count readings for signal 1 and signal 2.
The displayed values should be less than 7900 counts.
The dark current signal can be displayed directly on the screen.
1 Select in VWD - Settings - Analog Output as Raw Sample or Raw Reference.
2 Press the Plot button and select the Analog signal in Setup.
3 Return to the Plot screen by pressing Done.

DAC Test

This test determines correct operation of the digital-analog converter (DAC).
The test applies an AC voltage (10 µV) to the DC output of the DAC. The DAC output
is connected to the analog output connector at the rear of the detector.
68 1100 Series Variable Wavelength Detector Reference Manual