7
Control Module Screens for the Agilent 1100 Variable Wavelength Detector
Firmware Update
Use the Esc key to receive Views on the F5 key. Choose System from the pull-down
menu. Use the F3 key (Records) to select the VWD. Use the F5 key (FW Update) to
enter the Update section. If you want to update the resident firmware (together with
specific main firmware revisions), select the a file from the PCMCIA card
(RESnnnn.DLB) and press execute. If you want to update the main firmware, press F7
key (Transfer) to turn the module into the resident mode (LED on module should blink
yellow).
Use the Esc key to receive Views on the F5 key. Choose System from the pull-down
menu. Use the F3 key (Records) to select the Generic module. In this screen the
resident firmware revision is shown.
Use the F5 key (FW Update) to enter the Update section. Select the a file from the
PCMCIA card (1314nnnn.DLB) and press execute. When the update has finished, press