Figure 46 Block Diagram VWD Main Controller Functionality
Deuterium Lamp Filament Control
The deuterium lamp filament control circuit provides a constant voltage of 2.5 VDC at
approximately 6 A to the filament of the deuterium lamp. The deuterium lamp filament
1100 Series Variable Wavelength Detector Reference Manual 187