Agilent Technologies 1100 Series Reference Manual page 94

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Variable Wavelength Detector
Figure 19
Standard Flow Cell
Disassembling
the Flow Cell
CAUTION
Do not use tweezers to remove windows as the surfaces can easily be scratched.
Cleaning the Flow
Cell Parts
Always use new gaskets.
NOTE
Reassembling
the Flow Cell
The semimicro #1 and #2 gaskets (items 6 and 7,
NOTE
look
very similar. Do not mix them up.
Next steps
• Perform a leak test. If OK, insert the flow cell.
• Perform
92 1100 Series Variable Wavelength Detector Reference Manual
1 Unscrew the cell screw using a 4-mm hexagonal wrench.
2 Remove the SST rings using a pair of tweezers.
3 Use adhesive tape to remove the peek ring, the window and the gasket.
4 Repeat
step 1
through
otherwise they could be mixed !).
5 Pour isopropanol into the cell hole and wipe clean with a piece of lint-free cloth.
6 Clean the windows with ethanol or methanol. Dry it with a piece of lint-free cloth.
7 Hold the flow cell cassette horizontally and place gasket in position.
Ensure both cell holes can be seen through the holes of gasket.
8 Place the window on gasket.
9 Place the peek ring on the window.
10 Insert the conical springs. Make sure the conical springs point towards the window.
Otherwise tightening the cell screw might break the window.
11 Screw the cell screw into the flow cell and tighten the screw.
12 Repeat the procedure for the other cell side.
• Reconnect the capillaries, see
"Zero-Order Calibration"
on page 61 to check the correct positioning of the flow cell.
step 3
for the other window (keep the parts separate -
"Exchanging a Flow Cell"
on page 60 and
"Semimicro Flow Cell"
on page 91.
"656-nm Wavelength Calibration"
on page 149)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents