Installing the Autosampler

Flow Connections

Flow Connections

Parts required
#
1
Preparations
• Sampler is installed in the LC system
Toxic and hazardous solvents
WA R N I N G
The handling of solvents and reagents can hold health risks.
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
3 Connect the corrugated waste tube to the solvent waste from the leak plane.
4 Ensure that the waste tube is positioned inside the leak channel.
Do not extend the waste capillary of the autosampler. The siphoning effect might empty the
N O T E
complete seat capillary introducing air into the system.