6
Introduction to the Variable Wavelength Dector
Use the following tables for selecting the setting which you want to use for RS-232C
communication. The number 0 means that the switch is down and 1 means that the
switch is up.
Table 55
Table 56
Table 57
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.
202 1100 Series Variable Wavelength Detector Reference Manual
RS-232C
0
Baud Rate Settings
Switches
3
4
5
0
0
0
0
0
1
0
1
0
0
1
1
Data Bit Settings
Switch 6
Data Word Size
0
7 Bit Communication
1
8 Bit Communication
Parity Settings
Switches
Parity
7
8
0
0
No Parity
1
0
Odd Parity
1
1
Even Parity
1
Baud rate
Baud Rate
Switches
3
9600
1
1200
1
2400
1
4800
1
Data
Parity
Bits
Baud Rate
4
5
0
0
9600
0
1
14400
1
0
19200
1
1
38400

Hide quick links: