Response Of Outputs To Errors - Endress+Hauser Cerabar S PMP71 with MID Part Certificate Operating Instructions Manual

Process pressure measurement
Table of Contents

Advertisement

Troubleshooting
Output
A (Alarm)
Current output
– Device does not continue measuring.
– The current output assumes the value
specified via the OUTPUT FAIL MODE
ALT. CURR. OUTPUT
ALARM.
See also the following section: "Configuring
current output for an alarm".
Bar graph
The bar graph adopts the value defined by the
(onsite display)
OUTPUT FAIL MODE
Onsite display
– The measured value and message are
displayed alternately
– Measured value display:
permanently displayed.
Message display
– 3-digit number such as A122 and
description
Remote operation
In the case of an alarm, the ALARM STATUS
(FieldCare or HART
parameter displays a 3-digit number such as
handheld terminal)
122 for "Sensor not connected".
Menu path: (GROUP SELECTION →) OPERATING MENU → OUTPUT
1)
Menu path: (GROUP SELECTION →) OPERATING MENU → MESSAGES
2)
46
8.2

Response of outputs to errors

The device makes a distinction between the error types Alarm, Warning and Error.
See the following table and → ä 39, "Messages".
1)
,
1
and SET MAX.
1
1
parameter.
-symbol is
2)
W (Warning)
Device continues measuring.
The bar graph adopts the value which
corresponds to the current value.
– The measured value and message are
displayed alternately
– Measured value display:
-symbol
flashes.
Message display:
– 3-digit number such as W613 and
description
In the case of a warning, the ALARM
2
STATUS
parameter displays a 3-digit
number such as 613 for "Simulation is
active".
Cerabar S PMP71 with 4 to 20 mA HART
E (Error: Alarm/Warning)
For this error, you can enter whether the
device should react as in the event of an
alarm or as in the event of a warning.
See corresponding "Alarm" or "Warning"
column.
(See also Operating Instructions BA00413P,
parameter description SELECT ALARM
TYPE or these Operating Instructions)
See this table, "Alarm" or "Warning"
column, depending on the option selected.
– The measured value and message are
displayed alternately
– Measured value display: see
corresponding "Alarm" or "Warning"
column
Message display:
– 3-digit number such as E731 and
description
In the case of an error, the ALARM
2
STATUS
parameter displays a 3-digit
number such as 731 for "URV user limits
exceeded".
Endress+Hauser