Current Outputs - Endress+Hauser Smartec S CLD134 Operating Instructions Manual

Conductivity measuring system
Hide thumbs Also See for Smartec S CLD134:
Table of Contents

Advertisement

Smartec S CLD134
Coding
Field
Enter process
B4
temperature
Display temperature
B5
and calibrate
temperature sensor
Temperature
B6
difference is displayed
Coding
Field
Function group
O
OUTPUT
O1
Select current output
Enter linear
O2
O2 (1)
characteristic
O211 Select current range
0/4 mA value:
O212
enter corresponding
measured value
Endress+Hauser
Selection or range
(factory settings bold)
25 °C
-10.0 ... 150.0 °C
Display and entry of real
temperature
-35.0 ... 250.0 °C
0.0 °C
-5.0 ... 5.0 °C
6.4.3

Current outputs

The OUTPUT function group is used to configure the individual outputs.
Furthermore, a current output value can be simulated to check the current outputs (O2 (2)).
Selection or range
(factory settings bold)
out1
out2
lin = linear (1)
sim = simulation (2)
4 ... 20 mA
0 ... 20 mA
Cond: 0.00 μS/cm
Conc: 0.00 %
Temp.: -10.0 °C
entire measuring range
Display
°
C
25.0
B4
ProcTemp.
°
C
0.0
B5
RealTemp.
°
0.0
C
B6
TempOffs.
Display
O
OUTPUT
out1
O1
Sel. Out
lin
O2
Sel.Type
4-20
O211
Sel.Range
µS/cm
0
O212
0/4 mA
Commissioning
Info
Only if B1 = fixed.
This value can only be specified in °C.
This entry is used to calibrate the temperature
sensor to an external measurement.
Omitted if B1 = fixed.
The difference between the entered actual
value and the measured temperature is
displayed.
Omitted if B1 = fixed.
Info
Configuration of the current output (not
available for PROFIBUS versions).
A different characteristic can be selected for
each output.
The slope of the characteristic may be
positive or negative.
Enter the measured value corresponding to
the minimum current value (0/4 mA) at the
transmitter output.
Display format from A3.
(Spreading: see Technical data.)
41