Smartec CLD132/134 PROFIBUS PA/DP
Endress+Hauser
7

System integration

7.1

PROFIBUS PA/DP block model

In the PROFIBUS configuration, all the device parameters are categorized according to
their functional properties and tasks and are generally assigned to three different blocks.
A block may be regarded as a container in which parameters and the associated functions
are contained (see ).
A PROFIBUS device has the following block types:
• A physical block (device block)
The physical block contains all the device-specific features of the device.
• One or more transducer blocks
The transducer block contains all the measuring and device-specific parameters of the
device. The measuring principles (e.g. conductivity, temperature) are depicted in the
transducer blocks in accordance with the PROFIBUS Profile 3.0 Specification.
• One or more function blocks (function block)
A function block contains the device' s automation functions. The transmitter contains
Analog Input Blocks which can be used to scale the measured values and to check for
limit value overshoot.
A number of automation tasks can be implemented with these blocks. In addition to these
blocks, a transmitter can also contain any number of other blocks. These can include, for
example, several analog input function blocks if more than one process variable is provided
by the transmitter.
Local operation
Funktionsgruppe
Auswahl der
ALPHA-TABELLE
Tabellen
1
1 ... 4
T
(>1 nur bei MBU)
Funktionsgruppe
Auswahl der aktiven
KONZENTRATION
Konzentrationstabelle
; H SO ;
NaOH
2
4
H PO ; HNO
3
4
3
K
User
1
... 4
Funktionsgruppe
Auswahl der
Sprache
SERVICE
ENG
; GER
ITA; FRA
S
ESP; NEL
Sensor
Signal processing
 4
Block model (gray = profile blocks)
7.1.1
Physical block (device block)
A physical block contains all the data that uniquely identify and characterize the
transmitter. It is an electronic version of a nameplate on the transmitter. Physical blocks
parameters are e.g. device type, device name, manufacturer identification, serial number.
Auswahl der
Auswahl des
Eingabe des
Eingabe der Anzahl der
Tabellenoption
Tabellenstützpunkte
Tabellenwertepaares
Temperaturwertes
(x-Wert)
lesen
1
1
edit
1 ... 10
1 ... Anzahl aus R3
0,0 °C
T3
T1
T2
fertg
T4
–35,0 ... 250,0 °C
Multiplikationsfaktor
Auswahl der
Auswahl der
Eingabe der Anzahl der
für den Konzentrationswert
Tabellen
Tabellenoption
Tabellenstützpunkte
einer User-Tabelle
(nur bei MBU)
1
lesen
1
1
1 ... 4
edit
1 ... 16
K1
0,5 ... 1,5
K2
(>1 nur bei MBU)
K3
K4
Auswahl des
Eingabe des
HOLD-Konfiguration
Manueller HOLD
HOLD-Effektes
Festwertes
Kein = kein HOLD
(nur fall Fest)
S+C = bei Parametrieren
Letzt = letzter Wert
0
und Kalibrieren
aus
Setup = bei Parametrieren
Fest = fester Wert
0 ... 100 %
ein
CAL = bei Kalibrieren S4
S1
S2
von 20 bzw. 16 mA
S3
Manufacturer specific
parameters
Analyser Transducer Block 1
Main Meas. Value
Main Process Value
Analyser Transducer Block 2
Temperature Value
Temperature
System integration
T5
Physical Block
K5
S5
MRS
OUT VALUE
Analog Input Block 1
Value/Status
Main Meas. Value
Main Process Value
OUT VALUE
Analog Input Block 2
Value/Status
Temperature Value
Temperature
A0051925