Keysight N9911X User Manual
Keysight N9911X User Manual

Keysight N9911X User Manual

Economical waveguide calibration components
Table of Contents

Advertisement

Quick Links

A l l t e s t I n s t r u me n t s , I n c .
5 0 0 C e n t r a l A v e .
F a r mi n g d a l e , N J 0 7 7 2 7
P : ( 7 3 2 ) 9 1 9 - 3 3 3 9
F : ( 7 3 2 ) 9 1 9 - 3 3 3 2
a l l t e s t . n e t
s s a l e s @ a l l t e s t . n e t
T h e t e s t & me a s u r e me n t
e q u i p me n t y o u n e e d a t
t h e p r i c e y o u w a n t .
A l l t e s t c a r r i e s t h e w o r l d ' s l a r g e s t s e l e c t i o n o f
u s e d / r e f u r b i s h e d b e n c h t o p t e s t & me a s u r e me n t
e q u i p me n t a t 5 0 % t h e p r i c e o f n e w .
O O u r e q u i p me n t i s g u a r a n t e e d w o r k i n g , w a r r a n t i e d , a n d
a v a i l a b l e w i t h c e r t i f i e d c a l i b r a t i o n f r o m o u r i n - h o u s e s t a f f
o f t e c h n i c i a n s a n d e n g i n e e r s .
• 1 0 + f u l l t i me t e c h n i c i a n s w i t h o v e r 1 5 0 y e a r s o f
s p e c i a l i z a t i o n
• 9 0 d a y w a r r a n t y & 5 d a y r i g h t o f r e t u r n o n a l l
e q u i p me n t
• • 1 - 3 y e a r w a r r a n t i e s f o r n e w a n d
p r e mi u m- r e f u r b i s h e d e q u i p me n t
• E v e r y u n i t t e s t e d t o O E M s p e c i f i c a t i o n s
• S a t i s f a c t i o n g u a r a n t e e d
Y o u h a v e p l a n s , w e w i l l h e l p y o u a c h i e v e t h e m.
A n y p r o j e c t . A n y b u d g e t .
t
G e t a q u o t e t o d a y !
C C a l l ( 7 3 2 ) 9 1 9 - 3 3 3 9 o r e ma i l s a l e s @a l l t e s t . n e t .

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the N9911X and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight N9911X

  • Page 1 T h e t e s t & me a s u r e me n t e q u i p me n t y o u n e e d a t t h e p r i c e y o u w a n t . A l l t e s t I n s t r u me n t s , I n c .
  • Page 2 Keysight N9911X Economical Waveguide Calibration Components User’s Guide...
  • Page 4 User’s Guide N9911-90002...
  • Page 5 Should Keysight and the user have a separate written agreement with warranty terms covering the material WA R N I N G in this document that conflict with...
  • Page 6 To print copies of documentation from the Web, download the PDF file from the Keysight web site: • Go to http://www.keysight.com. • Enter the product model number (Ex: N9911X) in the Search bar. • Click Search. • Click the Manuals hyperlink.
  • Page 7 User’s Guide N9911-90002...
  • Page 8: Table Of Contents

    Returning a Component to Keysight ........
  • Page 9 Contents User’s Guide N9911-90002...
  • Page 10: General Information

    General Information User’s Guide N9911-90002 1- 1...
  • Page 11: Overview

    N9911X Economical Waveguide Calibration Components are used to calibrate FieldFox analyzers. With the calibration data properly loaded in the analyzer and a measurement calibration completed, systematic errors are minimized. For information on components available to purchase, refer to the online document Keysight N9911X Economical Waveguide Calibration Components Configuration Guide (part number N9911-90003).
  • Page 12: Recording The Components Serial Numbers

    The N9911X calibration components are manufactured by Flann Microwave. Each component is imprinted with a Flann logo, part ID number, and serial number. To determine the Keysight option number for a component, use the cross-references in the following tables.
  • Page 13 Component Keysight Flann Serial Number Option Number Part ID Number Calibration Components – Metric – 16091–NM70–6336 Adapter 1, waveguide to type-N (male) coax N9911X – 16091–NM70–6336 Adapter 2, waveguide to type-N (male) coax N9911X – 16045–6337 Termination N9911X – Flush short 16191–6338...
  • Page 14 Component Keysight Flann Serial Number Option Number Part ID Number Calibration Components – Metric – 18091–NM–6340 Adapter 1, waveguide to type-N (male) coax N9911X – 18091–NM–6340 Adapter 2, waveguide to type-N (male) coax N9911X – 18045–6341 Termination N9911X – Flush short 18191–6342...
  • Page 15 Keysight Flann Serial Number Option Number Part ID Number Calibration Components – Metric – 20091–JM–2371 Adapter 1, waveguide to 3.5 mm (male) coax N9911X – 20091–JM–2371 Adapter 2, waveguide to 3.5 mm (male) coax N9911X – 20045–2376 Termination N9911X –...
  • Page 16: Preventive Maintenance

    General Information Preventive Maintenance Preventive Maintenance The best techniques for maintaining the integrity of the components include: • routine visual inspection • routine cleaning • proper connection techniques All of these are described in Chapter 3. Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any component mated to it.
  • Page 17 General Information Preventive Maintenance 1- 8 User’s Guide N9911-90002...
  • Page 18: Specifications

    Specifications User’s Guide N9911-90002 2- 1...
  • Page 19: Environmental Requirements

    Specifications Environmental Requirements Environmental Requirements Table 2-1 Environmental Requirements Parameter Limits Temperature +4 °C to +40 °C Operating Storage -20 °C to +70 °C 1. The temperature range over which the calibration components maintain performance to their specifications. Temperature—What to Watch Out For Changes in temperature can affect electrical characteristics.
  • Page 20: Typical Vswr Values

    Specifications Typical VSWR Values Typical VSWR Values Table 2-2 Table 2-3 list the typical VSWR values for the N9911X terminations and waveguide-to-coax adapters. Table 2-2 Typical VSWR Values for Terminations Keysight Waveguide Band Designator Thread Frequency VSWR, Option Number Type...
  • Page 21: Instrument Interface Specifications

    Specifications Instrument Interface Specifications Instrument Interface Specifications Table 2-4 N9911X Instrument Interface Specifications Waveguide Thread Type Compatible Flange Type Band Designator Metric Flange (UAR 70, PAR 70, CAR 70, UDR 70, RDR 70, PDR 70) C-Band/ WR137/ English Flange (UG-441/U, UG-344/U, UG-343B/U, UG-440B/U, UG-1733/U, UG-1732/U,...
  • Page 22: Dimension Specifications

    Specifications Dimension Specifications Dimension Specifications Table 2-5 N9911X Dimension Specifications Component Parameter WR137 WR90 WR62 WR42 Adapter, Waveguide aperture 46 um MAX 25 um MAX 23 um MAX 20 um MAX waveguide-to-coax tolerance Termination Waveguide aperture 46 um MAX 25 um MAX...
  • Page 23 Specifications Dimension Specifications 2- 6 User’s Guide N9911-90002...
  • Page 24 Use, Maintenance, and Care of the Components User’s Guide N9911-90002 3- 1...
  • Page 25: Use, Maintenance, And Care Of The Components Electrostatic Discharge

    Use, Maintenance, and Care of the Components Electrostatic Discharge Electrostatic Discharge Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged.
  • Page 26: Visual Inspection

    Use, Maintenance, and Care of the Components Visual Inspection Visual Inspection Visual inspection and, if necessary, cleaning should be done every time a connection is made. Inspect mating surfaces for dirt, dust, foreign particles, or scratches, which can degrade component performance. A damaged mating surface can damage any good surface connected to it.
  • Page 27: Cleaning The Mating Plane Surfaces

    Use, Maintenance, and Care of the Components Cleaning the Mating Plane Surfaces Cleaning the Mating Plane Surfaces 1. Use Compressed Air or Nitrogen WARNING Always use protective eyewear when using compressed air or nitrogen. Use compressed air (or nitrogen) to loosen particles on the mating plane surfaces. Clean air cannot damage a component or leave particles or residues behind.
  • Page 28: Connections

    Use, Maintenance, and Care of the Components Connections Connections Good connections require a skilled operator. Slight errors in operator technique can have a significant effect on measurements and measurement uncertainties. The most common cause of measurement error is poor connections. The following procedures illustrate how to make good connections.
  • Page 29 Use, Maintenance, and Care of the Components Connections 1. Place the slip pins in the top and bottom holes of one flange, as shown in Figure 3-3. Figure 3-3 Inserting Slip Pins 2. Using the pins as guides for the adapter, offset shim, and waveguide-to-coax adapter, carefully align the flanges and insert two screws in the diagonal corner holes, as shown in Figure 3-4.
  • Page 30: Tightening A Flange Connection

    Use, Maintenance, and Care of the Components Connections Tightening a Flange Connection NOTE The best connection has symmetrical pressure applied as you gradually tighten the screws. 1. In an “X” pattern (for equal compression), tighten all four screws using a hex ball driver. Do not over-tighten.
  • Page 31: Connecting A Termination To A Waveguide-To-Coax Adapter

    Use, Maintenance, and Care of the Components Connections Connecting a Termination to a Waveguide-to-Coax Adapter Figure 3-6 Termination and Adapter Connecting an Offset Shim Between a Flush Short and Waveguide-to-Coax Adapter Create an offset short by connecting the offset shim between the short and the appropriate adapter, as shown in Figure 3-7.
  • Page 32: Connecting A Flush Short To A Waveguide-To-Coax Adapter

    Use, Maintenance, and Care of the Components Connections Connecting a Flush Short to a Waveguide-to-Coax Adapter Figure 3-8 Flush Short and Adapter User’s Guide N9911-90002...
  • Page 33: Handling And Storage

    Use, Maintenance, and Care of the Components Handling and Storage Handling and Storage • Install the protective end caps and store the calibration components when not in use. • Never store components loose in a box, or in a desk or bench drawer. This is the most common cause of component damage during storage.
  • Page 34: Troubleshooting

    Troubleshooting User’s Guide N9911-90002 4- 1...
  • Page 35: Troubleshooting Process

    Troubleshooting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration, or if your analyzer does not pass performance verification, follow the steps Figure 4-1 Figure 4-1 Troubleshooting Flowchart 4- 2 User’s Guide N9911-90002...
  • Page 36: Where To Look For More Information

    If you need additional information, see “Contacting Keysight” on page 4-4. Returning a Component to Keysight If an N9911X component requires service, contact Keysight Technologies for information on where to send it - see “Contacting Keysight” on page 4-4. Please provide the following information: •...
  • Page 37: Contacting Keysight

    Troubleshooting Contacting Keysight Contacting Keysight Assistance with test and measurement needs and information on finding a local Keysight office are available on the Web at: www.keysight.com/find/assist If you do not have access to the Internet, please contact your Keysight field engineer.
  • Page 38: Component Dimensions

    Component Dimensions User’s Guide N9911-90002 5- 1...
  • Page 39 “Adapter, Waveguide Designators C-Band/WR137/WG14, Metric, Keysight Part Number N9911X-110” on page 5-10 “Adapter, Waveguide Designators C-Band/WR137/WG14, English (Imperial), Keysight Part Number N9911X-115” on page 5-11 “Adapter, Waveguide Designators X-Band/WR90/WG16, Metric, Keysight Part Number N9911X-210” on page 5-12 “Adapter, Waveguide Designators X-Band/WR90/WG16, English (Imperial), Keysight Part Number N9911X-215”...
  • Page 40 Table 5-1 List of Figures Figure Title and Location “Termination, Waveguide Designators C-Band/WR137/WG14, English (Imperial), Keysight Part Number N9911X-116” on page 5-19 “Termination, Waveguide Designators X-Band/WR90/WG16, Metric, Keysight Part Number N9911X-211” on page 5-20 “Termination, Waveguide Designators X-Band/WR90/WG16, English (Imperial), Keysight Part Number N9911X-216”...
  • Page 41 Table 5-1 List of Figures Figure Title and Location “Flush Short, Waveguide Designators C-Band/WR137/WG14, English (Imperial), Keysight Part Number N9911X-117” on page 5-35 “Flush Short, Waveguide Designators X-Band/WR90/WG16, Metric, Keysight Part Number N9911X-212” on page 5-36 “Flush Short, Waveguide Designators X-Band/WR90/WG16, English (Imperial), Keysight Part Number N9911X-217”...
  • Page 42: Flange Dimensions

    Component Dimensions Flange Dimensions Flange Dimensions NOTE The dimensions shown in the following graphics are in millimeters. Figure 5-1 Flange, Waveguide Designators C-Band/WR137/WG14, Metric User’s Guide N9911-90002...
  • Page 43: Flange, Waveguide Designators C-Band/Wr137/Wg14, English (Imperial)

    Component Dimensions Flange Dimensions Figure 5-2 Flange, Waveguide Designators C-Band/WR137/WG14, English (Imperial) Figure 5-3 Flange, Waveguide Designators X-Band/WR90/WG16, Metric 5- 6 User’s Guide N9911-90002...
  • Page 44: Flange, Waveguide Designators X-Band/Wr90/Wg16, English (Imperial)

    Component Dimensions Flange Dimensions Figure 5-4 Flange, Waveguide Designators X-Band/WR90/WG16, English (Imperial) Figure 5-5 Flange, Waveguide Designators Ku-Band/WR62/WG18, Metric User’s Guide N9911-90002...
  • Page 45: Flange, Waveguide Designators Ku-Band/Wr62/Wg18, English (Imperial)

    Component Dimensions Flange Dimensions Figure 5-6 Flange, Waveguide Designators Ku-Band/WR62/WG18, English (Imperial) Figure 5-7 Flange, Waveguide Designators K-Band/WR42/WG20, Metric 5- 8 User’s Guide N9911-90002...
  • Page 46: Flange, Waveguide Designators K-Band/Wr42/Wg20, English (Imperial)

    Component Dimensions Flange Dimensions Figure 5-8 Flange, Waveguide Designators K-Band/WR42/WG20, English (Imperial) User’s Guide N9911-90002...
  • Page 47: Waveguide-To-Coax Adapter Dimensions

    Component Dimensions Waveguide-to-Coax Adapter Dimensions Waveguide-to-Coax Adapter Dimensions NOTE The dimensions shown in the following graphics are in millimeters. Figure 5-9 Adapter, Waveguide Designators C-Band/WR137/WG14, Metric, Keysight Part Number N9911X-110 5- 10 User’s Guide N9911-90002...
  • Page 48 Component Dimensions Waveguide-to-Coax Adapter Dimensions Figure 5-10 Adapter, Waveguide Designators C-Band/WR137/WG14, English (Imperial), Keysight Part Number N9911X-115 User’s Guide N9911-90002 5-11...
  • Page 49 Component Dimensions Waveguide-to-Coax Adapter Dimensions Figure 5-11 Adapter, Waveguide Designators X-Band/WR90/WG16, Metric, Keysight Part Number N9911X-210 5- 12 User’s Guide N9911-90002...
  • Page 50 Component Dimensions Waveguide-to-Coax Adapter Dimensions Figure 5-12 Adapter, Waveguide Designators X-Band/WR90/WG16, English (Imperial), Keysight Part Number N9911X-215 User’s Guide N9911-90002 5-13...
  • Page 51 Component Dimensions Waveguide-to-Coax Adapter Dimensions Figure 5-13 Adapter, Waveguide Designators Ku-Band/WR62/WG18, Metric, Keysight Part Number N9911X-310 5- 14 User’s Guide N9911-90002...
  • Page 52 Component Dimensions Waveguide-to-Coax Adapter Dimensions Figure 5-14 Adapter, Waveguide Designators Ku-Band/WR62/WG18, English (Imperial), Keysight Part Number N9911X-315 User’s Guide N9911-90002 5-15...
  • Page 53 Component Dimensions Waveguide-to-Coax Adapter Dimensions Figure 5-15 Adapter, Waveguide Designators K-Band/WR42/WG20, Metric, Keysight Part Number N9911X-410 5- 16 User’s Guide N9911-90002...
  • Page 54 Component Dimensions Waveguide-to-Coax Adapter Dimensions Figure 5-16 Adapter, Waveguide Designators K-Band/WR42/WG20, English (Imperial), Keysight Part Number N9911X-415 User’s Guide N9911-90002 5-17...
  • Page 55: Termination Dimensions

    Component Dimensions Termination Dimensions Termination Dimensions NOTE The dimensions shown in the following graphics are in millimeters. Figure 5-17 Termination, Waveguide Designators C-Band/WR137/WG14, Metric, Keysight Part Number N9911X-111 5- 18 User’s Guide N9911-90002...
  • Page 56: Termination, Waveguide Designators C-Band/Wr137/Wg14, English (Imperial)

    Component Dimensions Termination Dimensions Figure 5-18 Termination, Waveguide Designators C-Band/WR137/WG14, English (Imperial), Keysight Part Number N9911X-116 User’s Guide N9911-90002 5-19...
  • Page 57: Termination, Waveguide Designators X-Band/Wr90/Wg16, Metric, Keysight Part

    Component Dimensions Termination Dimensions Figure 5-19 Termination, Waveguide Designators X-Band/WR90/WG16, Metric, Keysight Part Number N9911X-211 5- 20 User’s Guide N9911-90002...
  • Page 58: Termination, Waveguide Designators X-Band/Wr90/Wg16, English (Imperial), Keysight Part Number N9911X-216

    Component Dimensions Termination Dimensions Figure 5-20 Termination, Waveguide Designators X-Band/WR90/WG16, English (Imperial), Keysight Part Number N9911X-216 User’s Guide N9911-90002 5-21...
  • Page 59: Termination, Waveguide Designators Ku-Band/Wr62/Wg18, Metric, Keysight Part Number N9911X-311

    Component Dimensions Termination Dimensions Figure 5-21 Termination, Waveguide Designators Ku-Band/WR62/WG18, Metric, Keysight Part Number N9911X-311 5- 22 User’s Guide N9911-90002...
  • Page 60: Termination, Waveguide Designators Ku-Band/Wr62/Wg18, English (Imperial)

    Component Dimensions Termination Dimensions Figure 5-22 Termination, Waveguide Designators Ku-Band/WR62/WG18, English (Imperial), Keysight Part Number N9911X-316 User’s Guide N9911-90002 5-23...
  • Page 61: Termination, Waveguide Designators K-Band/Wr42/Wg20, Metric, Keysight Part

    Component Dimensions Termination Dimensions Figure 5-23 Termination, Waveguide Designators K-Band/WR42/WG20, Metric, Keysight Part Number N9911X-411 5- 24 User’s Guide N9911-90002...
  • Page 62: Termination, Waveguide Designators K-Band/Wr42/Wg20, English (Imperial), Keysight Part Number N9911X-416

    Component Dimensions Termination Dimensions Figure 5-24 Termination, Waveguide Designators K-Band/WR42/WG20, English (Imperial), Keysight Part Number N9911X-416 User’s Guide N9911-90002 5-25...
  • Page 63: Offset Shim Dimensions

    Component Dimensions Offset Shim Dimensions Offset Shim Dimensions NOTE The dimensions shown in the following graphics are in millimeters. Figure 5-25 Offset Shim, Waveguide Designators C-Band/WR137/WG14, Metric, Keysight Part Number N9911X-113 5- 26 User’s Guide N9911-90002...
  • Page 64: Offset Shim, Waveguide Designators C-Band/Wr137/Wg14, English (Imperial), Keysight Part Number N9911X-118

    Component Dimensions Offset Shim Dimensions Figure 5-26 Offset Shim, Waveguide Designators C-Band/WR137/WG14, English (Imperial), Keysight Part Number N9911X-118 User’s Guide N9911-90002 5-27...
  • Page 65: Offset Shim, Waveguide Designators X-Band/Wr90/Wg16, Metric, Keysight Part

    Component Dimensions Offset Shim Dimensions Figure 5-27 Offset Shim, Waveguide Designators X-Band/WR90/WG16, Metric, Keysight Part Number N9911X-213 5- 28 User’s Guide N9911-90002...
  • Page 66: Offset Shim, Waveguide Designators X-Band/Wr90/Wg16, English (Imperial), Keysight Part Number N9911X-218

    Component Dimensions Offset Shim Dimensions Figure 5-28 Offset Shim, Waveguide Designators X-Band/WR90/WG16, English (Imperial), Keysight Part Number N9911X-218 User’s Guide N9911-90002 5-29...
  • Page 67: Offset Shim, Waveguide Designators Ku-Band/Wr62/Wg18, Metric, Keysight Part Number N9911X-313

    Component Dimensions Offset Shim Dimensions Figure 5-29 Offset Shim, Waveguide Designators Ku-Band/WR62/WG18, Metric, Keysight Part Number N9911X-313 5- 30 User’s Guide N9911-90002...
  • Page 68: Offset Shim, Waveguide Designators Ku-Band/Wr62/Wg18, English (Imperial)

    Component Dimensions Offset Shim Dimensions Figure 5-30 Offset Shim, Waveguide Designators Ku-Band/WR62/WG18, English (Imperial), Keysight Part Number N9911X-318 User’s Guide N9911-90002 5-31...
  • Page 69: Offset Shim, Waveguide Designators K-Band/Wr42/Wg20, Metric, Keysight Part

    Component Dimensions Offset Shim Dimensions Figure 5-31 Offset Shim, Waveguide Designators K-Band/WR42/WG20, Metric, Keysight Part Number N9911X-413 5- 32 User’s Guide N9911-90002...
  • Page 70 Component Dimensions Offset Shim Dimensions Figure 5-32 Offset Shim, Waveguide Designators K-Band/WR42/WG20, English (Imperial), Keysight Part Number N9911X-418 User’s Guide N9911-90002 5-33...
  • Page 71: Flush Short Dimensions

    Component Dimensions Flush Short Dimensions Flush Short Dimensions NOTE The dimensions shown in the following graphics are in millimeters. Figure 5-33 Flush Short, Waveguide Designators C-Band/WR137/WG14, Metric, Keysight Part Number N9911X-112 5- 34 User’s Guide N9911-90002...
  • Page 72: Flush Short, Waveguide Designators C-Band/Wr137/Wg14, English (Imperial), Keysight Part Number N9911X-117

    Component Dimensions Flush Short Dimensions Figure 5-34 Flush Short, Waveguide Designators C-Band/WR137/WG14, English (Imperial), Keysight Part Number N9911X-117 User’s Guide N9911-90002 5-35...
  • Page 73: Flush Short, Waveguide Designators X-Band/Wr90/Wg16, Metric, Keysight Part

    Component Dimensions Flush Short Dimensions Figure 5-35 Flush Short, Waveguide Designators X-Band/WR90/WG16, Metric, Keysight Part Number N9911X-212 5- 36 User’s Guide N9911-90002...
  • Page 74: Flush Short, Waveguide Designators X-Band/Wr90/Wg16, English (Imperial), Keysight Part Number N9911X-217

    Component Dimensions Flush Short Dimensions Figure 5-36 Flush Short, Waveguide Designators X-Band/WR90/WG16, English (Imperial), Keysight Part Number N9911X-217 User’s Guide N9911-90002 5-37...
  • Page 75: Flush Short, Waveguide Designators Ku-Band/Wr62/Wg18, Metric, Keysight Part Number N9911X-312

    Component Dimensions Flush Short Dimensions Figure 5-37 Flush Short, Waveguide Designators Ku-Band/WR62/WG18, Metric, Keysight Part Number N9911X-312 5- 38 User’s Guide N9911-90002...
  • Page 76: Flush Short, Waveguide Designators Ku-Band/Wr62/Wg18, English (Imperial), Keysight Part Number N9911X-317

    Component Dimensions Flush Short Dimensions Figure 5-38 Flush Short, Waveguide Designators Ku-Band/WR62/WG18, English (Imperial), Keysight Part Number N9911X-317 User’s Guide N9911-90002 5-39...
  • Page 77: Flush Short, Waveguide Designators K-Band/Wr42/Wg20, Metric, Keysight Part

    Component Dimensions Flush Short Dimensions Figure 5-39 Flush Short, Waveguide Designators K-Band/WR42/WG20, Metric, Keysight Part Number N9911X-412 5- 40 User’s Guide N9911-90002...
  • Page 78: Flush Short, Waveguide Designators K-Band/Wr42/Wg20, English (Imperial), Keysight Part Number N9911X-417

    Component Dimensions Flush Short Dimensions Figure 5-40 Flush Short, Waveguide Designators K-Band/WR42/WG20, English (Imperial), Keysight Part Number N9911X-417 User’s Guide N9911-90002 5-41...
  • Page 79 Component Dimensions Flush Short Dimensions 5- 42 User’s Guide N9911-90002...
  • Page 80 VSWR values requirements general information visual inspection environmental compressed air temperature for cleaning return component to Keysight connecting return loss specification handling 3-10 flanges how often to calibrate flush short to adapter how to calibrate for ESD protection...
  • Page 81 Index dimension electrical frequency instrument interface return loss static electricity storage 3-10 table mat, for ESD protection temperature cautions about requirements termination dimensions 5-18 troubleshooting, flowchart typical VSWR values component visual inspection VSWR typical values when to calibrate the analyzer wrist strap, for ESD protection Index-2 User’s Guide N9911-90002A...