CT Analyzer User Manual
6.2

CT-Object Card

6.2.1

Available Soft Keys

80
The CT-Object card is the most important card and always required for all types
of tests. In this card all necessary settings for a test are done. Some of the
settings contained in the CT-Object card are also shown in other cards for
information.
Note: The CT-Object card is the top user interface level. Pressing the
several times will always bring you back to the CT-Object card.
Figure 6-1
CT-Object card with default settings after initializing a new CT
test
Clears the results of the previous test and enables to start a new
test. All measurement results as well as all the parameters
previously determined using the "guesser function" are cleared. Text
entries for "Location" and "Object" remain unchanged.
Saves the test results and test settings to the currently loaded
CT Analyzer report file. If the test results have not been saved yet,
the Save As dialog is opened. Saving a file is described on page 64.
Saves the test results and test settings to a new .xml file. Saving a
file is described on page 64.
Allows to load a test stored on the Compact Flash card in order to
check its results, to recalculate the results with different settings or to
use its settings for a new test.
Note: Recalculation of the existing CT test results in order to verify
the behavior of the CT at different burdens or primary currents can be
performed by changing the burden value on the CT-Object card or
the primary current on the Ratio card. The recalculated results can
then be stored again like for a regular test.
key