Test Settings And Results; Running A Residual Magnetism Measurement - Omicron Lab CT Analyzer User Manual

Table of Contents

Advertisement

6.4.1

Test Settings and Results

The following settings and test results are displayed in the Residual Magnetism
card.
6.4.2

Running a Residual Magnetism Measurement

1. Press the Select Cards soft key in the CT-Object card to open the Select
2. If not already specified, enter the I
Parameter
Description
I-sn
Nominal secondary current of the CT as entered in the
CT-Object card.
The I
sn
of the residual magnetism measurement. Otherwise, a
corresponding message is displayed.
Residual Flux
Absolute value [in Vs] of the residual magnetic flux in the CT
determined by the CT Analyzer.
The residual flux is always displayed after measurement,
even if the CT Analyzer could not determine the residual
magnetism.
Residual
Residual magnetism [in %] of the CT, calculated from the
Magnetism
residual flux and the maximum flux of the CT.
No value is displayed if the CT Analyzer was not able to
reach the knee point of the excitation curve.
Cards page. Enable the Residual Magnetism test card in the Select Cards
page. Press the Back soft key to apply your test card selection and return to
the CT-Object card.
Figure 6-7
Select Cards page with Residual Magnetism test card
enabled
of the CT has to be specified prior to the execution
of the CT on the CT-Object card.
sn
The CT Analyzer Test Cards
103