930.xxx
Error [930.xxx] Measurement input Vsec defective, contact next Omicron service
center.
Reason:
Solution:
931.xxx
Error [931.xxx] Measurement input Vprim defective, contact next Omicron service
center.
Reason:
Solution:
932.xxx
Error [932.xxx] Data error.
Reason:
Solution:
933.xxx
Error [933.xxx] Hardware failure, contact next Omicron service center.
Reason:
Solution:
Safety action: Power supply is switched off and all measurements are disabled.
934.xxx
Warning [934.xxx] Reverse power, don't disconnect any wires, don't switch off mains,
wait until power is dissipated. I = xxxxA.
Reason:
Solution:
Safety action: All measurements are stopped until reverse power is dissipated.
935.xxx
Error [935.xxx] No valid CT Analyzer software. Insert a CF-Card with valid Software and
press "Update Firmw.".
Reason:
Solution:
936.xxx
Warning [936.xxx] Corrupt Calibration Data for voltage inputs. Change to tools menu
and try to reset factory calibration. Until update of calibration data the device will be not
calibrated.
Reason:
Solution:
937.xxx
Warning [937.xxx] Corrupt Calibration Data for power output. Change to tools menu
and try to update factory calibration. Until update of calibration data the device will be
not calibrated.
Reason:
Solution:
938.xxx
Warning [938.xxx] MIF data block 1 corrupt.
Reason:
Solution:
939.xxx
Warning [939.xxx] MIF data block 2 corrupt.
Reason:
Solution:
Power supply for analog input "Sec" is defective.
Contact your next OMICRON service center.
Power supply for analog input "Prim" is defective.
Contact your next OMICRON service center.
The sequence of reading of the analog input channels is not correct.
If this error occurs more frequently, contact your next OMICRON service
center.
Temperature detection circuit on secondary side is defective.
Send the device to your next OMICRON service center.
The device receives reverse power of > 20mA.
Wait until the power is dissipated within the device and the error message
disappears.
Cannot find valid CT Analyzer software.
Insert a Compact Flash card with valid software (CTAnalyzer.bin) in the
directory A:\Omicron.
The calibration data checksum for the analog inputs is not correct.
Try to update the factory calibration using the tools menu. If this does not solve
the problem, contact your next OMICRON service center.
The calibration data checksum for the power output is not correct.
Try to update the factory calibration using the tools menu. If this does not solve
the problem, contact your next OMICRON service center.
The checksum of the factory settings data is faulty.
Licensing or device settings data are possibly corrupt. Contact your next
OMICRON service center.
The checksum of the factory settings data is faulty.
Licensing or device settings data are possibly corrupt. Contact your next
OMICRON service center.
Error and Warning Messages
183