CT Analyzer User Manual
940.xxx
941.xxx
942.xxx
943.xxx
944.xxx
184
Warning [940.xxx] CMOS data block corrupt. Check all device settings. Press Clear
Values to work with default values. Press OK to work with corrupt device settings.
Reason:
Device settings data block corrupt.
Solution:
Check all device settings.
Error [941.xxx] Corrupt factory calibration data for voltage inputs. Contact next
Omicron service center.
Reason:
Factory calibration data block for voltage inputs corrupt.
Solution:
Contact your next OMICRON service center.
Error [942.xxx] Corrupt factory calibration data for power unit. Contact next Omicron
service center.
Reason:
Factory calibration data corrupt.
Solution:
Contact your next OMICRON service center.
Error [943.xxx] Error overwriting calibration data.
Reason:
Factory calibration data cannot be restored.
Solution:
Contact your next OMICRON service center.
Warning [944.xxx] Corrupt user text! Change to tools menu and try to update text. Until
update of user text, the device will use default text.
Reason:
Checksum for user language support data invalid.
Solution:
Copy a user text file (CTUser_xxx.bin) to the directory A:\Omicron\ on the
Compact Flash card and try to update the user language support data using the
"Update Text" function in the tools menu.