Omicron Lab CT Analyzer User Manual page 131

Table of Contents

Advertisement

Param.
Description
Class
Accuracy class according to standard.
Δϕ
Phase deviation.
ε
Current ratio error.
FS
Instrument security factor (direct measurement
method).
FSi
Instrument security factor (indirect meas. method).
Parameters assessed for IEEE C57.13
Parameter
Description
Class
Accuracy class according to
standard.
Rct
DC winding resistance.
Δϕ
Phase deviation.
RCF
Ratio correction factor.
ε
Current ratio error at secondary
@ Isn
current I
.
sn
ε
@ 20 * Isn Current ratio error at 20 times the
secondary current I
Vb
Rated secondary terminal voltage.
Vknee
Knee point voltage.
Vk / Ik
User-defined measuring point.
Vk1 / Ik1
User-defined measuring point 1.
1. Acc. to IEEE C57.13 (1993)
The CT Analyzer Test Cards
Parameter assessed for
IEEE C57.13
protection CTs,
class
C
T
x
x
x
x
x
x
.
sn
x
x
x
x
Parameter assessed
for IEC 61869-2
metering CTs
0.1, 0.2, 0.2s,
0.5, 0.5s, 1
3, 5
x
x
x
x
x
x
x
x
x
IEEE
C57.13.6
meter.
high
CTs
accuracy
metering
CTs
0.15,
0.3, 0.6,
1
0.15s
X
K
1.2
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
129