Omicron Lab CT Analyzer User Manual page 167

Table of Contents

Advertisement

Values without the prefix "!" in the ratio table of the Ratio card have guaranteed
accuracy. The accuracy of values marked with a "!" in the table is reduced by
factor 2 since these values are not directly measured but calculated from the
measured values instead.
Ratio measurement accuracy for 1 A CTs at rated current
CT ratio
I
sn
0.2 - 1
1
> 1 - 2000
1
> 2000 - 5000
1
> 5000 - 10000
1
1. Nominal burden of the CT.
Ratio measurement accuracy for 5 A CTs at rated current
CT ratio
I
sn
0.2 - 1
5
> 1 - 2000
5
> 2000 - 5000
5
> 5000 - 10000
5
1. Nominal burden of the CT.
Phase measurement accuracy at rated current
Resolution
ϕ
Accuracy (cos
0.8 - 1)
Turns ratio measurement accuracy
Resolution
Accuracy
1
Rated power
Typical
accuracy
1.0 - 30VA
0.05%
0 - 30VA
0.02%
0 - 30VA
0.03%
0 - 30VA
0.05%
1
Rated power
Typical
accuracy
1.0 - 75VA
0.05%
0 - 75VA
0.02%
0 - 75VA
0.03%
0 - 75VA
0.05%
0.01min
1min (typical)
3min (guaranteed)
0.01 turns
0.05% (typical)
0.1% (guaranteed)
Technical Data
Guaranteed
accuracy
0.1%
0.05%
0.1%
0.2%
Guaranteed
accuracy
0.1%
0.05%
0.1%
0.2%
165