CT Analyzer User Manual
905.xxx
906.xxx
908.xxx
911.xxx
912.xxx
929.xxx
182
Error [905.xxx] Over-temperature of power supply, wait until device has cooled down.
Reason:
Secondary side power supply has reached the warning temperature limit and
entered save mode to cool down.
Solution:
When this error is acknowledged as long it is active, the popup window is
closed and only the message in the status line remains active until the
overtemperature disappears. If the error is not acknowledged, the popup
window remains active.
Comment:
The status line displays the flashing message "905.xxx Overtemp.".
Safety action: All measurements are stopped and the error message is displayed in the status
line and a popup window. After acknowledgement, the software waits at least
1 minute before the test can be started again.
Error [906.xxx] Excess temperature of power supply, wait until device has cooled
down.
Reason:
The primary side power supply has reached its warning temperature limit and
entered save mode to cool down.
Solution:
When this error is acknowledged as long it is active, the popup window is
closed and only the message in the status line remains active until the
overtemperature disappears. If the error is not acknowledged, the popup
window remains active.
Comment:
The status line displays the flashing message "906.xxx Overtemp.".
Safety action: All measurements are stopped and the error message is displayed in the status
line and a popup window. After acknowledgement, the software waits at least
1 minute before the test can be started again.
Error [908.xxx] Device shut down.
Reason:
Internal power failure on the measurement interface module.
Solution:
The message is active as long as the internal power for the measurement
interface fails. Contact the next OMICRON service center if the error does not
disappear.
Error [911.xxx] Power supply error, contact next Omicron service center. Desired
voltage xxxx, measured voltage yyyy.
Reason:
One supply voltage on the measurement interface boards is out of tolerance.
Solution:
Contact your next OMICRON service center.
Comment:
It is no longer possible to work with the device.
Error [912.xxx] Excess internal temperature, wait until device has cooled down.
Reason:
Device-internal temperature at measurement interface > 75°C.
Solution:
Excessive heating of the device. Prevent the device from direct sunlight and
wait until it has cooled down. If the error occurs without previous exposure to
direct sunlight, the error is probably caused by a hardware error. Please
contact your next OMICRON service center.
Comment:
The status line displays the flashing message "912.xxx Overtemp.".
Safety action: All measurements are stopped and the error is displayed in the status line and
in a popup window. The error is active until the temperature falls below 60°C.
Error [929.xxx] Hardware failure, contact next Omicron service center.
Reason:
The circuit for reverse power dissipation does not work properly.
Solution:
Contact your next OMICRON service center.
Safety action: All measurements are stopped and the power supply is switched off.