Omicron Lab CT Analyzer User Manual page 179

Table of Contents

Advertisement

415.xxx
Warning [415.xxx] Quick measurement cannot be started. Please disconnect the
CT SB2 switch box.
Reason:
Solution:
501.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected
to the CT Analyzer.
Reason:
Solution:
504.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected
to the CT Analyzer.
Reason:
Solution:
505.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected
to the CT Analyzer.
Reason:
Solution:
510.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected
to the CT Analyzer.
Reason:
Solution:
511.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected
to the CT Analyzer.
Reason:
Solution:
513.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected
to the CT Analyzer.
Reason:
Solution:
514.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected
to the CT Analyzer.
Reason:
Solution:
800.xxx
Warning [800.xxx] Flash card access error. Internal file system error.
Reason:
Solution:
801.xxx
Warning [801.xxx] Flash card access error. Invalid file name or path.
Reason:
Solution:
You are trying to initiate a Quick measurement with the CT SB2 switch box still
connected to the CT Analyzer.
Measurements using Quick Test are not possible when the CT SB2 is
connected to the CT Analyzer. Remove all connections to the CT SB2 and
restart your Quick measurement on the CT Analyzer.
The remote interface handler couldn't decipher the command.
Remote interface error. Check the connection between CT Analyzer and PC.
Data transmission error.
Check the connection between CT Analyzer and PC.
Data transmission error.
Check the connection between CT Analyzer and PC.
Remote interface error: Failed to get the required measurement data because
the measurement is still running.
Check the connection between CT Analyzer and PC.
Remote interface error: Failed to unzip the transmitted data block.
Check the connection between CT Analyzer and PC.
The current remote interface command is not allowed.
Make sure that the current CT Analyzer state allows the execution of the
command.
Remote interface error: Remote command was submitted with an invalid
parameter.
Verify the parameter (index) submitted with the remote interface command.
The file system sent an error message.
Compact Flash card is possibly corrupt. Use a new Compact Flash card.
The file system sent an error message.
The Compact Flash card is possibly corrupt. Try to save the data to a PC and
format the CF card.
Error and Warning Messages
177

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ct sb2

Table of Contents