Error [824.xxx] An internal software error has occurred at address xxxxxxxH. To log
this error, make sure that a CF card is inserted and then press OK. Please send the file
\OMICRON\ErrorLog.xml on the CF card to the next OMICRON service center.
Reason:
Invalid result in a floating point operation (NaN).
Solution:
Contact your next OMICRON service center.
Warning [825.xxx] Size of file [FileName] >= 1Mbyte.
Reason:
The size of the file copied to the CT Analyzer Compact Flash card or virtual disc
is > 1 MB.
Solution:
Avoid transfer of large files > 1 MB to the CT Analyzer.
License < ... > is invalid.
Reason:
An invalid license entry was found in the license file Omicron.lic on the CF
Card.
Solution:
Specify a valid license key.
Setting value for < ... > is invalid. The default value of < ... > is used.
Reason:
Invalid item in file Settings.inf.
Solution:
The file Settings.inf in the Omicron directory on the CF card of the
CT Analyzer has a formatting error in one of the parameters. Delete this file and
check the settings in the menu Main Menu -> Settings -> Misc. Settings. If
necessary, change the settings and store the file again.
Lifetime of relays near end. Contact the next OMICRON service center soon to arrange
a hardware service.
Reason:
The relays inside the CT Analyzer near the end of their lifetime (more than
1,000,000 measurements performed).
Solution:
Arrange a hardware service in order to replace the relays.
You are about to load a report with an invalid hash code. Some measurement values
may not be authentic.
Reason:
The file checksum is not valid.
Solution:
Check the authenticity of the report file loaded.
No valid license was found in this file!
Reason:
No valid license found.
Solution:
Check the license file. The license possibly does not correspond to the serial
number of the CT Analyzer.
Downgrade a firmware below version 3.53 is not supported.
Reason:
Attempt for a downgrade to a version number < 3.53.
Solution:
Make sure that die FW version number for the upgrade/downgrade is not below
3.53.
Error [900.xxx] Power supply error. Switch off the device and wait 1 minute to restart.
Reason:
The output power was larger than 350VA or the power supply is defective.
Solution:
Switch the CT Analyzer off, wait 1 minute and then switch it on again. If this
error occurs more frequently, you should contact your next OMICRON service
center.