CT Analyzer User Manual
10.8.2
Shock and Vibration
10.8.3
Mechanical Data
10.8.4
Safety Standards, Electromagnetic Compatibility (EMC)
168
Dynamics
Vibration
Shock
Weight, Dimensions and Protection
Weight
Dimensions W x H x D
CE Conformity, Requirements
The product adheres to the specifications of the guidelines of the Council of
the European Community for meeting the requirements of the member states
regarding the electromagnetic compatibility (EMC) Directive 2004/108/EC and
the low-voltage Directive 2006/95/EC.
EMC
Emission
Europe
International
USA
Immunity
Europe
International
Certified Safety Standards
Europe
International
USA
Tested according to IEC 60068-2-6;
frequency range 10 ... 150 Hz; acceleration
2
2g continuous (20 m/s
Tested according to IEC 60068-2-27
(operating mode); 15g / 11ms, half-sinusoid,
3 shocks in each axis
< 8kg (17.6lbs) without accessories
360 x 285 x 145mm
(14.2 x 11.2 x 5.7")
EN 61326-1 Class A
IEC 61326-1 Class A
FCC Subpart B of Part 15 Class A
EN 61326-1
IEC 61326-1
EN 61010-1
IEC 61010-1
UL 61010-1
); 20 cycles per axis