Climate
Operating temperature
Storage and transportation
Max. altitude
Humidity
Dynamics
Vibration
Shock
Weight, Dimensions and Protection
Weight
Dimensions W x H x D
Degree of protection
–10 ... +50 °C (14 ... 122 °F)
–20 ... +70 °C (–4 ... 158 °F)
2000m
5 ... 95% relative humidity, non-condensing
Tested acc. to IEC 60068-2-78, Cab, Damp
Heat: Temp. 40°C, duration 48h,
rel. humidity 95%
Tested according to IEC 60068-2-6;
frequency range 10 ... 150 Hz; acceleration
2
2g continuous (20 m/s
Tested according to IEC 60068-2-27
(operating mode); 15g / 11ms, half-sinusoid,
3 shocks in each axis
2.6kg (5.7lbs) without accessories
285 x 68 x 225mm
(11.2 x 2.7 x 8.9")
IP20 according to EN 60529
Technical Data
); 20 cycles per axis