6.8

Assessment Card

The Assessment card is only available if it is enabled in the Select Cards page
(Default Test Card Selection or Select Cards soft key in the CT-Object card).
Depending on the standard and the type of CT (protection or metering CT), the
according parameters are listed.
The column for automatic assessment ("Auto") is automatically filled after the
test is finished. The following assessments are possible:
It is also possible to perform a manual assessment for the individual parameters.
To do this, select the parameter to be assessed using the
and apply your assessment using the OK, Failed or ? soft key.
Figure 6-19
Note: Automatic assessment is only performed for the CT behavior at nominal
burden ("VA" parameter on the CT-Object card). For the CT behavior at
operating burden ("Burden" parameter on the CT-Object card), no automatic
assessment is performed.
"OK": The results measured for this parameter comply with the requirements
defined by the selected standard and the parameters in the CT-Object card.
"Failed": The results do not comply with the requirements.
"n/a": No assessment possible due to one of the following reasons:
Comparison with input parameter is not possible.
Assessment does not make sense due to incorrect polarity or invalid
measurement value.
Assessment card (example)