CT Analyzer User Manual
98
Specific parameters and settings displayed for IEEE C57.13
protection CTs
The following parameter is only displayed in the CT-Object card if the
IEEE C57.13 standard is selected with the type "Prot. CT".
Param.
Description
Vb *
Rated secondary terminal voltage.
If the IEEE C57.13 standard is selected with the type
"protection CT" (class C, K or T), the VA and Cos
parameters are not accessible to the user. In this case, the
user has to enter the terminal voltage V
Possible values: Integer values from 10 to 1200V or soft
2
keys ?
, 10V, 20V, 50V, 100V, 200V, 400V or 800V.
Default: ? or ?*.
* This parameter is relevant for assessment. No automatic
assessment is available if the user did not enter data for
this parameter prior to the test (see page 82).
Vk
User-defined measuring point.
User-defined measuring point 1.
Vk1
Possible values: Integer values from 0.1V to 20.000V or
1
soft key ?
.
Default: ?.
Ik
User-defined measuring point.
User-defined measuring point 1.
Ik1
Possible values: Integer values from 0.03mA to
30.000mA or soft key ?
Default: ?.
RE(20*Isn) Ratio error at 20 * I
Possible values: Integer values from 1% to 99%.
Default: 10%.
Rct
Specified secondary winding resistance.
Possible values: 0 to 3000 ohms or soft key ?
Default: ?
1. Acc. to IEEE C57.13 (1993)
2. If the question mark is entered for this parameter and a new test is started, the CT Analyzer
automatically tries to determine the value using its integrated guesser function (see chapter 7 on
page 131).