Leica tps1100 Reference Manual page 81

Hide thumbs Also See for tps1100:
Table of Contents

Advertisement

GeoCOM Reference Manual
AUT_RC_MOTOR_ERROR
RC_FATAL
RC_ABORT
AUT_RC_NO_TARGET
AUT_RC_MULTIPLE_TARGETS
AUT_RC_BAD_ENVIRONMENT
AUT_RC_DEV_ERROR
AUT_RC_NOT_ENABLED
AUT_RC_DETECTOR_ERROR
RC_COM_TIMEDOUT
See Also
AUS_SetUserAtrState
AUS_GetUserAtrState
AUT_SetFineAdjustMode
AUT_GetFineAdjustMode
Example
RC_TYPE
ON_OFF_TYPE ATRState;
double
dHzSearchRange=0.08;// search range in [rad]
dVSearchRange=0.08; // search range in [rad]
Result = AUS_GetUserAtrState(ATRState);// The ATR-
if(ATRState==ON)
{
// performs a fine position with a max. target
// search range of 0.08rad (5gon) in Hz and V
// direction
TPS1100 - Version 1.05
Result;
dHzSearchRange, dVSearchRange
Instrument has no 'motorization'.
Fatal error.
Function aborted.
No target found.
Multiple targets found.
Inadequate environment conditions.
During the determination of the angle
deviation error detected, repeat fine
positioning
ATR mode not enabled, enable ATR
mode
AZE error, at repeated occur call service
Communication time out. (perhaps
increase COM timeout, see
COM_SetTimeout
// Status must
// be set for
// the fine
// adjust
// functionality
Automation - AUT
)
7-24

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents