6.2 Replacement Level Supported

The replacement level supported for this product is to the printed circuit board (PCB)
and major subassembly level. Once you isolate a suspected PCB, follow the procedures
in Disassembly Guide, to replace the PCB with a known good PCB. Check to see if the
trouble symptom persists, swap back the replacement PCB with the suspected
malfunctioning PCB (the original PCB that was installed when you started trouble
shooting) and continue troubleshooting as detected in this section.

6.3 Prior to Disassembly

1. Turn the N5500 off by pressing the Power On/Off switch.
2. Disconnect the monitor from the AC power source

6.4 Fuse Replacement

This section describes the fuse replacement without disassembling the main case of the
monitor.
AC Inlet Fuse Replacement
N5500 Service Manual
Disassembly Guide
a. Push the fuse socket latch of the AC inlet
slightly down.
b. Pull the socket out from the inlet.
47