Power-On Settings and Service Functions
Reset Softkey
Save Softkey

Print Softkey Menu

34
.
The RESET softkey can be used if any settings stored in memory have been
changed from factory default values. If YES is pressed, the instrument sounds
three tones and the settings return to factory default values. When NO is
pressed, no changes are made to the settings stored in memory.
When adjustable values are changed from factory default, the SAVE softkey
can be used to preserve the settings as institutional power-on default values.
Pressing YES stores the current settings in memory. The instrument sounds
three tones indicating that the changes have been saved as power-on default
values. The new saved values will continue to be used through power-on and
off cycles until they are changed and saved again, or until they are reset. If NO
is pressed, the changed values will not be saved.
Note:
An invalid tone indicates that a parameter value cannot be saved as a
power-on default. See Power-On Settings on page 29. Along with the
invalid tone, a message will be displayed indicating which parameter
could not be saved as a power-on default.
Accessing the PRINT softkey makes four printouts available. See Data Port
Interface Protocol on page 89, for information about how to make
connections to the data port and how data is presented in a printout. The
appropriate printout can be selected by pressing the corresponding softkey.
The softkey configuration that appears after the PRINT softkey has been
pressed is shown below.
Up to 48 hours of trend data can be viewed on the printouts described below.
When the monitor is turned on, trend data is recorded every 4 seconds. As an

Hide quick links: