GE Healthcare Brivo XR118 Service Manual page 44

Table of Contents

Advertisement

GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
Table 4-1 Target System Requirements
No.
10
11
3.) The item values can be measured on target system or use the certificates/x-ray quality
4.) Purchaser needs to make sure target system's exposure function works well before installation
Note:
Note:
Note:
The recommended grid type is 70LP/CM.
Target System Grid Type
28 LP/CM
30 LP/CM
32 LP/CM
34 LP/CM
36 LP/CM
38 LP/CM
40 LP/CM
42 LP/CM
44 LP/CM
46 LP/CM
48 LP/CM
50 LP/CM
52 LP/CM
54 LP/CM
56 LP/CM
58 LP/CM
Page 44
6
EVISION
Categor
Item
y
System
Regulatory
System
Radiation Protection
(accuracy) inspection and calibration reports periodically issued to target system as reference.
of Brivo XR118. Please make sure the image quality of analog film/CR/digital image acquired
from target system can meet customer's clinical requirement.
If the grid type of target system is not in the list above (36LP/CM, 40LP/CM, 60LP/CM, 70LP/
CM, 78LP/CM), it's recommended to configure the "Utility-->Image Processing-->Grid line
rate" for GLR algorithm according to the table below.
If the effect of grid line reduction is not obvious or the image is distorted after the image
reprocessed using the configuration shown as below, then it's recommended to change the
grid.
For following scenario, please turn off the GLR algorithm in "Utility-->Image Processing-->Grid
Line Reduction Single Energy".
-
Bucky (vibration grid)
-
No grid
Preferred Capability
Shall comply with local medical
equipment regulatory
requirement, such as GB9706/
IECv2/IECv3
Comply with GB9706.12-1997/
IEC601-1-3:1994 or IEC 60601-
1-3: 2008.
Recommend Grid Type for GLR
78LP/CM
70LP/CM
70LP/CM
36LP/CM
36LP/CM
36LP/CM, 40LP/CM
40LP/CM
40LP/CM
40LP/CM
40LP/CM
40LP/CM
40LP/CM, 60LP/CM
60LP/CM
60LP/CM
60LP/CM
60LP/CM
-
B
XR118 S
M
RIVO
ERVICE
ANUAL
Target
System
Claim
Pass
/Fail