4. Upon releasing the associated detector successfully, associated detector
will be updated as non-associated.
 If the detector release failed, the associated detector will not become
Detector Sleeping Setting
The detector sleep setting window displays the current sleep time of the
associated detector.
Note:
If the detector sleeping time is set to 30 minutes, the detector it will
automatically go into the sleeping mode after the system in 30 minutes idle
mode. Once the detector is in sleeping mode you must press the power button on
the detector for one to two seconds to exit the sleep mode. The detector will
wake up and be ready for an exposure in 10 seconds.
Detector Auto-Sensing Setting
The detector autosensing setting window displays the current autosensing time
of the associated detector.
Note:
Page 272
6
EVISION
non-associated. A popup message will be displayed to indicate user of the
release failure.
Figure 12-16
System - Detector Management- Detector Sleep Time
Detector sleep time value means the timeout value from detector's idle mode to sleep mode.
Figure 12-17
System - Detector Management - Detector Auto-sensing Time
Detector auto-sensing time means the auto-sensing timeout value from detector entering
auto-sensing mode.
Chapter 12 - - Set Preferences