Chapter 5 - Installation & Configuration
Section 1.0

Before You Begin

Before installation work begins, the site should be checked to verify the correctness of junction box
sizes, their locations, the number and sizes of interconnecting wiring and proper power availability
in the room.
Three standard boxes will be delivered: Main box, monitor and detector box.
If optional parts ordered, there will be a fourth box.
6
EVISION
Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
The distribution center or factory must be notified immediately of any damage or shortage of
parts.
Main Box
Detector Box
Chapter 5 - - Installation & Configuration
Monitor Box
Optional Box
Figure 5-2 Boxes Received