GE Healthcare Brivo XR118 Service Manual page 142

Table of Contents

Advertisement

GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
Bad line example:
After manual BP, take an extra exposure to confirm the specified bad line disappear.
5.7
Detector Replacement
If detector has too many bad pixels, redo calibration again, if still fail, suggest to replace detector.