Section 2.0

Digital Detector

The device or portions of the device may have power present from the batteries even when shut
down, please demount the battery first.
Detector Assembly
Before the replacement, make sure detector's malfunction is not caused by collision.
1.) Open the service port at the back of detector.
2.) Check the vibration sensor.
3.) After the detector assembly replaced, please refer to
6
EVISION
If the center of the vibration sensor is Transparent, it means no any collision during usage.
If the center of the vibration sensor is RED, it means collision exists during usage. Please
inform the user to carry the detector carefully.
Normal
register and associate the new detector with system.
Chapter 10 - - Renewal parts
B
XR118 S
RIVO
ERVICE
Collision
Detector Association on page 117
M
ANUAL
to