The EMI Reduction algorithm has a maximum allowable correction. Any correction [amplitude
since the correction removes sinusoids] to the image will be smaller than or equal to this value.
The EMI Reduction algorithm is an iterative process. One of the stopping criteria is based on
the maximum amplitude of the suspected signal. In order to further control the algorithm (and
the associated computation time), the maximum number of iterations can be set to end the
iterative loop before the amplitude threshold is reached.
EMI Reduction parameters should be only changed adapt to site EMI circumstance.
This function need Service Key CD (Class M) in the DVD Drive to activate.