18.) How to perform XR118 Workstation IP configuration
19.) Image quantity displays wrong
6
EVISION
with superuser.
ii.)
Press keyboard "i" to enable edit function, add a line: route add -host "IP of pacs" gw "ip
of gateway".
iii.) Press ESC key and then input :wq to save and exit from configuration file.
iv.) Reboot system.
Solution:
-
eth0: IP: 3.1.1.2. Internal IP to connect with detector.
-
eth1: External IP to connect with hospital network, IP need be set in the same segment
as hospital RIS/PACS/Printer IP
-
Lo: loop IP, IP: 127.0.0.1.
Problem:
Although images are deleted, but the image quantity still shows big number in Image
Management window.
Solution:
i.)
Go to Transfer log and cleanup the transfer log by deleting all failed/queues/pending jobs
ii.)
Press Ctrl+Shift+F5 to open ENGGUI and open command line window
iii.) Type command line: cd /database/DicomImage
iv.) Type command line: for f in *; do sudo rm "$f"; done
v.)
Restart the application
vi.) Expected outcome: Extra occupied disk should be freed up.