Section 9.0

Printer Calibration

The following applies when performing Printer calibrations.
Note:
Only performing printer calibration when printer has problem.
Follow the instructions presented on SUIF.
Section 10.0

PACS / Image Display Test Image

10.1
Objectives
10.2
Simplified Procedure
1.) Verify All PACS Review Monitors Are Calibrated
Note:
2.) Send PACS Test Images to PACS
Page 148
6
EVISION
Simplified Procedure: To determine configuration for Burn-on-Send or No-Burn for each
destination host (PACS).
Results: To record results from Simplified Procedure and Additional Tests.
Background: To provide further information about Image Display on PACS.
Additional Tests: To test several Image Display characteristics (Optional).
Monitor calibration is a critical first step before testing and configuring Image Display.
a.) Verify that the PACS monitor is calibrated according to DICOM Part 14. Each site is
responsible for providing evidence of PACS monitor calibration.
b.) View SMPTE Pattern
i.) Load SMPTE pattern into Patient List by clicking on "Image Tools" then "SMPTE"
ii.) Send SMPTE pattern to PACS review workstation.
iii.) Verify that 5% and 95% squares are equally visible on PACS.
c.)
Record results in
Table
Section 9.0 - Printer Calibration
6-2, line numbers 1 and 2.