12.1.1
Personnel Requirements
Personnel Requirements
1
12.1.2
Overview
This procedure can be performed with:
• Fluke tester
12.1.3
Tools and Test Equipment
Item
Fluke ESA612 or equivalent
Note:
The equivalent means that the tool is designed to be capable of performing tests to both the
standard of IEC 62353 and the earth leakage current of IEC 60601-1.
RISK OF ELECTRIC SHOCK.
VOLTAGE PRESENT.
THE SYSTEM MUST BE DISCONNECTED FROM THE FACILITY MAINS INCLUDING THE A1
PANEL GROUND WIRE.
12.1.4
Setup the tools
Refer to Chapter 1
12.1.5
Leakage Current Test Procedure
• Earth Leakage Current Test
No
1
1) Connected the BRIVO XR118 power plug to tool output receptacle.
Page 128
6
EVISION
Preliminary Reqs
Not Applicable
4.5.5 Set up tools on page 29
TEST POINT
Connected the BRIVO XR118
power plug to tool output
receptacle
Section 12.0 - Leakage Current Test
Procedure
30min
Qty.
Effectivity
Part#
1
-
Expected Value
(mA)
10
B
XR118 S
RIVO
ERVICE
Finalization
10min
Manufacturer
-
Fluke
Measure Value
RESULT
(mA)
M
ANUAL