8.4
Restore
8.4.1
Before You Begin
Note:
The detector calibration data table_IDC can not be restored in different system.
Note:
If you wish to use a USB flash drive, select [USB DRIVE] and install drive into USB port.
8.4.2
Restore System Files
1.) From the application desktop, click on the Utilities icon.
2.) Click on [Launch] button.
3.) Select the Utilities Tab,
4.) Click on the plus sign preceding the System folder ICON, in the navigation frame.
5.) Click on [Restore] link,
6.) Insert your "Cal/Config Backup" CD/DVD into the DVD drive.
6
EVISION
After the OS Configuration restore is run, the SUIF will no longer respond because of the files
that have been changed. A system Reset must be performed after restoring the OS
Configuration files to continue using the SUIF.
You must have a "good" backup of the files you wish to restore from CD/DVD.
The "Backup" media can only be inserted after the restore utility screen is displayed.