5461425-8EN, R
IRECTION
9.8
Operator Manual (OM) Installation
The operator manual must be installed:
The procedure follows:
1.) Verify that no archival process is running.
2.) Verify that system is not in a state where Service Browser is invoked.
3.) Press the following keys: <Cntrl><Alt><F2> simultaneously, to open a command window.
4.) At the login prompt, type: xruser
5.) Enter the password: 4$xray
6.) Insert Operators manual CD into the tray of DVD drive PC
7.) To run the inter-active script, at prompt type the following: ./xrdinstall.om, click ENTER key.
8.) A message is displayed: Do you want to proceed? Enter "y", click ENTER key.
9.) Below message is displayed:
10.) A message is displayed: Installing Online Manual in all languages.
6
EVISION
-
when the system is initially installed
-
following a complete system load from cold
-
following a software reload of the workstation PC
Select the language to install:
1 for English
2 for French
3 for Spanish
4 for Chinese
5 for Turkish
6 for Indonesian
7 for Russian
8 for Portuguese
9 for Korean
10 for All Languages
11 for Quit, no installation shall happen
Enter "10" for Enter Choice. Click ENTER key.
Installation Complete will be displayed after OM installation.