7.) Set exposure techniques as: SID 150cm; 60kV; 5mAs.
8.) Take exposure 5 times for each exposure mode (Detector under the table and above the
9.) After exposures done, view Raw Image with cobb tool at same location to compare the
10.) Compare the average values under the same kV to make sure there is no uniformity issue
11.) For the wall stand, follow the same procedure.
12.) Please contact the target system supplier if the uniformity issue is caused by the table top, wall
Section 4.0

Acquisition Test

1.) Create a new patient. Select "chest" exam, AEC mode. Insert detector into detector housing.
2.) Press hand switch do an exposure.
3.) Close the exam.
Section 5.0

Regulatory Test

Brivo XR118 system meets all related requirements of IEC v3. Brivo XR118 is an upgrade kit and
has no x-ray generating mechanism, so only image processing related regulatory tests are
applicable (x-ray generating related regulatory tests are not applicable).
Page 162
6
EVISION
table).
average value.
caused by the table top.
stand front panel or unsteady exposure.
Align the tube to detector. Open a suitable FOV. Check if there no exposure inhibit displayed.
Chapter 7 - - Brivo XR118 System Check
B
XR118 S
M
RIVO
ERVICE
ANUAL