GE Healthcare Brivo XR118 Service Manual page 220

Table of Contents

Advertisement

GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
Solutions:
1.) Perform Manual bad pixel calibration.
2.) Take exposure again to confirm the bad lines disappear.
3.) Perform QAP again.
5.2
Equi-spaced Zebra White Bad Lines Issue
Detector calibration passed but QAP failed, after taking an exposure, Equi-spaced zebra white bad
lines appeared in whole image.
Page 220
6
EVISION
QAP Pass
Issued Image
Issued Image
Section 5.0 - Image Quality Issues
B
See Section
4.0.