Table 12-8 Preference Tab
Option
1
2
3
4
5
6
7
8
9
6
EVISION
Option 5 - Checking this option will cause the workstation PC to strip out all patient information
in the images being sent to the remote node being configured. Same as de-identify and push
automatically. Some installations may want to check this option for nodes being configured for
security purposes. For example, (if you do not want to divulge any patient information to the
people using this remote node).
Options 6 - should be checked / enabled.
Options 7 - should be selected if the network host has been configured to accept MPPS
messages from the magic.
Options 8 - Should be selected if the network host is HIS/RIS system.
Options 9 - If selected Image will be sent to this network node as CR format.
Allow this host to query the (system name)
Allow this host to retrieve from the (system name)
Allow this host to send images to the (system name)
This network host accepts multiple frame image.
Perform de-identification (anonymous patient images)
when sending to this network host.
Apply "Burn-On-Send" to images when sending to this
network host. See PACs calibration for information.
(PACS Calibration & Test Image)
Send MPPS N-Create and N-Set notification to this
network host.
Make this host the HIS/RIS source. (Only one host can
be designated as the HIS/RIS source.)
DX Images sent as CR Modality
Chapter 5 - - Installation & Configuration