GE Healthcare Brivo XR118 Service Manual page 30

Table of Contents

Advertisement

GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
• Power on the Tool
- Press the power switch on the left-side panel so the "I" side of the ac power switch is depressed.
The tool will perform a series of self tests and then display the message shown as below picture
when the self test has completed successfully.
• Setting the test standard to IEC 62353
- The tool is designed to perform electrical safety testing based on a number of different safety
standards: AAMI ES1/NFPA99, IEC62353, IEC60601-1, and AN/NZS 3551. AAMI is the Analyzer's
default standard.
- To select the IEC 62353 standard:
1.) Press "Setup".
2.) From the setup menu, press "F4" button under the softkey labeled "More" to reveal additional
menu selections.
3.) Press "F2" button under the softkey labeled "Instrument" to reveal the instrument setup
selections.
4.) Press "F1" button under the softkey labeled "Standard" to open the scroll box above the softkey
label.
5.) Press button of UP or Down to scroll through the standard selections.
6.) When the desired standard is displayed, press "F1" button under the softkey labeled
"Standard" to confirm the standard.
7.) Press "F4" button under the softkey labeled "Done" to completed the standard setup.
• Accessing the tool's Leakage current test interface
1.) Press "POINT TO POINT" button to access below interface
2.) Press "F3" button under the softkey labeled "Leakage" to access Leakage current interface.
3.) Make zero the test lead resistance. Connected the two alligator clips accessory as below. Press
"F4" button under the softkey labeled "Zero Leads". The tool zeroes out the measurement to make
zero the test lead resistance.
4.5.6
Protective Earth Resistance Test
1. Shut down the system.
Page 30
6
EVISION
Section 4.0 - Equipment Safety - Electrostatic Discharge (ESD)
B
XR118 S
M
RIVO
ERVICE
ANUAL