Root Cause:
Press exposure switch too quickly after exposure count down clock is displayed in screen
during Detector calibration, so detector scan data is lost.
Solutions:
1.) Do not press exposure switch immediately after exposure count down clock is displayed in
2.) Repeat Detector calibration
3.) Repeat QAP.
6
EVISION
screen. You can delay a little seconds to take exposure, only if you can finish the exposure
within the time.