Detector Registration
A detector must be registered with an x-ray system at the first time when the two work together. The
registration is accomplished by connecting the detector and the 118 system through the registration
process.
Note:
The active Digital Detector is identified by an asterisk (*).
Note:
After registered and specified as an associated detector in the system, about 3~5 minutes of
association time is required to upload its calibration file to system.
10.2.1
Adding A New Detector Identified By Name
1.) Press the [Add] button on the Detector Registration screen.
2.) Enter new detector's MAC address.
Note:
3.) Choose Name for the Select Identification Type.
Page 118
6
EVISION
Figure 5-16 Add a detector By Name
Please detach the detector battery to find the mac address.
Section 10.0 - Detector Association
Figure 5-15 Detector Registration
B
XR118 S
M
RIVO
ERVICE
ANUAL