Additional Tests
The following tests are meant to provide additional information about PACS behavior and
performance. These tests should be used if further troubleshooting is required beyond the
configuration determined in the "Simplified Procedure".
Topics include
1.) Send All PACS Test Images
2.) DICOM Headers
6
EVISION
Send All PACS Test Images.
DICOM Headers.
Visible WC and WW Values. Image Saturation.
Reduced Dynamic Range with Burn-on-Send.
Interactive Windowing.
Annotations.
Multiple VOI-LUTs.
CR-fallback.
Send all images to PACS according to the following rules. Send images 1 through 5 as No-
Burn, and send images 6 through 8 as Burn-On-Send. Finally, if the site is interested in
exploring CR-Fallback as an option, configured the PACS for CR-fallback with help from the
local IT team. Then send image 9 as No-Burn.
a.) View Header
Open the PACS Test Image(s) on the PACS (Patient ID 2000002)
View the DICOM header information - The DICOM header is often viewable
Chapter 6 - - Calibration
B
XR118 S
M
RIVO
ERVICE
ANUAL
Page 155