General Cleaning and Painting
Keep all surfaces clean.
Remove power from the system before beginning work.
Check that power is off.
1.) Remove power from the system.
2.) Carefully clean all surfaces of the system.
3.) Re-apply power to the system.
3.1.2
Visual Inspection
Check for debris that would indicate abnormal wear.
Remove power from the system before beginning work.
Check that power is off.
1.) Remove power from the system.
2.) If any debris, wear of damage is observed, replace or repair the affected areas.
3.) Reapply power to the system.
6
EVISION
Pay attention to any paint damage when you clean the system.
Never use anything other than soap and water to clean plastic surfaces. Other cleaners may
damage the plastic.
Never use any corrosive, solvent or abrasive detergents or polishes.
Ensure that no water or other liquid can enter any equipment. This precaution prevents short
circuits and corrosion forming on components.
Methods of disinfection used must conform to legal regulations and guidelines regarding
disinfection and explosion protection.
If disinfectants are used which form explosive mixtures of gases, these gases must have
evaporated before switching on the equipment again.
Disinfection by spraying is not recommended because the disinfectant may enter the X-ray
equipment.
If room disinfection is done with an atomizer, it is recommended that the equipment be
switched OFF, allowed to cool down and covered with a plastic sheet. When the disinfectant
mist has subsided, the plastic sheet may be removed and the equipment be disinfected by
wiping.