Adding A New Detector Identified By Color and Shape
1. Press the [Add] button on the Detector Registration screen.
2. Enter the mac address of the detector you wish to register to the system.
 Please detach the detector battery to find the mac address.
3. Chose "Color and Shape"for the Select Identification Type.
4. Chose the Color and Shape from the pulldown menus.
5. Chose to make the new Digital Detector "Associated"or
"Non-Associated".
 If the detector is registered as associated, there will be a message
 If the association is successful, the detector will be marked as
 If the association failed, the detector information will be stored in the
6. Press [Save]. The new Digital Detector is now registered with the system.
7. Attach the plastic tag with the appropriate shape and color onto the Digital
Detector.
6
EVISION
Figure 12-14
System - Detector Management- Add By Color and Shape
window for user to confirm that "Detector association status would
change for already present associated detector" if there is already an
associated detector registered to the system.
associated detector with an asterisk (*) prefixed to the detector name in
the detector registration window. Any previous associated detector in the
system will be automatically changed to non-associated.
registered detector list and will not be marked as associated detector..
Chapter 12 - - Set Preferences
B
XR118 S
M
RIVO
ERVICE
ANUAL
Page 269